HP 1660E Series Service Manual page 128

Logic analyzers
Table of Contents

Advertisement

Testing Performance
Performance Test Record (oscilloscope)
Performance Test Record (oscilloscope)
Test
Sellings
Results
Vultage
Limits
Measured
Measurement
Accuracy
Channell
Zero Input
-13.7 V to -14.3 V
-3.43 V to -3.57 V
-1.37 V to -1.43 V
-137.0 mV to -143.0 mV
143.0 mVto 137.0 mV
1.43 V to 1.37 V
3.57 V to 3.43 V
14.3Vto 13.7 V
Channel 2
Zero Input
-13.7 Vto -14.3 V
-3.43 Vto -3.57 V
-1.37Vto-l.43V
-137.0 mVto -143.0 mV
143.0 mV to 137.0 mV
1.43 V to 1.37 V
3.57 V to 3.43 V
14.3Vto 13.7V
Offset Accuracy
Zero-input offset
Channell
4 VlOlv
0.00 V±320.0 mV
1VlOlv
0.00 V±80.0 mV
100 mVlOiv
0.00 V±8.0 mV
10 mVlOiv
0.00 V±800.0
fl V
Channel 2
4 VlOlv
0.00 V±320.0 mV
1VlOlv
0.00 V±80.0 mV
100 mVlOiv
0.00 V±8.0 mV
10 mVlOiv
0.00V±800.0 flV
DC-input offset
Channell
-35.00 V
-35.4 V to -34.6 V
-10.00 V
-10.1 Vto -9.90 V
-2.00 V
-2.02 V to -1.98 V
+2.00 V
+1.98 V to +2.02 V
+10.00 V
+9.90Vto+l0.1 V
+35.00 V
+34.6 V to +35.4 V
Channel 2
-35.00 V
-35.4 V to -34.6 V
-10.00 V
-10.1 Vto -9.90 V
-2.00 V
-2.02 V to -1.98 V
+2.00 V
+1.98 V to +2.02 V
+10.00V
+9.90Vto+l0.1 V
+35.00 V
+34.6 V to +35.4 V
3-102

Advertisement

Table of Contents
loading

This manual is also suitable for:

1660es series1660ep series