Testing Performance
Performance Test Record (oscilloscope)
Performance Test Record (oscilloscope)
Test
Sellings
Results
Vultage
Limits
Measured
Measurement
Accuracy
Channell
Zero Input
-13.7 V to -14.3 V
-3.43 V to -3.57 V
-1.37 V to -1.43 V
-137.0 mV to -143.0 mV
143.0 mVto 137.0 mV
1.43 V to 1.37 V
3.57 V to 3.43 V
14.3Vto 13.7 V
Channel 2
Zero Input
-13.7 Vto -14.3 V
-3.43 Vto -3.57 V
-1.37Vto-l.43V
-137.0 mVto -143.0 mV
143.0 mV to 137.0 mV
1.43 V to 1.37 V
3.57 V to 3.43 V
14.3Vto 13.7V
Offset Accuracy
Zero-input offset
Channell
4 VlOlv
0.00 V±320.0 mV
1VlOlv
0.00 V±80.0 mV
100 mVlOiv
0.00 V±8.0 mV
10 mVlOiv
0.00 V±800.0
fl V
Channel 2
4 VlOlv
0.00 V±320.0 mV
1VlOlv
0.00 V±80.0 mV
100 mVlOiv
0.00 V±8.0 mV
10 mVlOiv
0.00V±800.0 flV
DC-input offset
Channell
-35.00 V
-35.4 V to -34.6 V
-10.00 V
-10.1 Vto -9.90 V
-2.00 V
-2.02 V to -1.98 V
+2.00 V
+1.98 V to +2.02 V
+10.00 V
+9.90Vto+l0.1 V
+35.00 V
+34.6 V to +35.4 V
Channel 2
-35.00 V
-35.4 V to -34.6 V
-10.00 V
-10.1 Vto -9.90 V
-2.00 V
-2.02 V to -1.98 V
+2.00 V
+1.98 V to +2.02 V
+10.00V
+9.90Vto+l0.1 V
+35.00 V
+34.6 V to +35.4 V
3-102