To test the trigger sensitivity (oscilloscope)
Testing the trigger sensitivity verifies the performance of the following specifications:
• DC to 50 MHz: 0.063 x full scale (0.25 division)
• 50 MHz to 500 MHz: 0.125 x full scale (0.5 division)
Equipment Required
Equipment
Signal Generator
Cable
Adapter
Critical Specifications
50 MHz and 225 MHz, 30 - 80 mV RMS output
BNC 48-inch
Type N
(m)
to BNC
(f)
Recommended
ModeljPart
HP 8656B Opt 001
HP 8120-1840
HP 1250-0780
Set up the equipment
Tum on the equipment required and the logic analyzer. Let them warm up for 30
minutes
if
you have not already done so.