Set Up The Logic Analyzer - HP 1660E Series Service Manual

Logic analyzers
Table of Contents

Advertisement

Testing Performance
To tesllhe time interval accuracy (logic analyzer)
3 Set up the function generator according to the following table.
Function Generator Setup
Freq: 40.000 00 MHz
Amptd: 1.00 V
Modulation: Off
Set up the logic analyzer
1 Set up the Configuration menu.
a Press the Config key.
b
In
the Configuration menu, assign Pod 1 to Machine 1. To assign Pod 1, select the
Pod 1 field, then select Machine 1.
c
In
the Analyzer 1 box, select the Type field, then select TImtng.
~(
Conflgurollon
Anoly::er 1
Nome' (MACHWE I
)
Typ<,'
(~
Anol,per 2
Type'
~
3-61