To test the multiple-clock, multiple-edge, state acquisition
(logic analyzer)
Testing the multiple-clock, multiple-edge, state acquisition verifies the performance
of the following specifications:
• Minimum master to master clock time
• Maximum state acquisition speed
• SetuplHold time for multiple-clock, multiple-edge, state acquisition
• Minimum clock pulse width
This test checks two combinations of data using multiple clocks at two selected
setup/hold times.
Equipment Required
Equipment
Pulse Generator
Digitizing Oscilloscope
Adapter
SMA Coax Cable ((lty 31
Coupler
BNC Test Connector,
6><2 (Qty41
Critical Specifications
100 MHz 3.5 ns pulse width, < 600 ps rise time
;;, 6 GHz bandwidth, < 58 ps rise time
SMA(m)-BNC(fl
18 GHz bandwidth
BNC(mllml
Recommended
ModellPall
HP 8133A option 003
HP 54750A, with HP 54751 A
plugin
HP 1250-1200
HP 8120-4948
HP 1250-0216
Set up the equipment
1 Turn on the equipment required and the logic analyzer. Let them warm up for
30 minutes before beginning the test if you have not already done so..
2 Set up the pulse generator.
a Set up the pulse generator according to the following table.
Pulse Generator Setup
Timebase
Mode:
Int
Period: 10.000 ns
Channel 2
Mode: Pulse
Divide: PULSE ... 2
Width: 4.500 ns
High: -0.90 V
Low: -1.70V
Period
Divide: Divide ... 2
Ampl: 0.50 V
Dffs: 0.00 V
Channell
Mode: Pulse
Delay: 0.000 ns
Width: 3.500 ns
High: -0.90 V
Low: -1.70 V
b Disable the pulse generator channell COMP (with the LED oft).