Operating the Alaris® GW Volumetric Pump
Alaris® GW Volumetric Pump
Introduction & Start up
Follow the instructions supplied with the individual IV infusion set.
Use of non-specified IV infusion sets may impair the operation of the pump and the
accuracy of the infusion.
Refer to the "Compatible IV Infusion Sets" section in Appendix D, for a list of
recommended IV infusion sets.
Ensure that the tubing is inserted completely into the pumping channel, avoiding any
slack.
Position the IV fluid container to avoid spillage onto the pump.
When using 273-003 IV infusion sets, ensure a separation of at least 30cm is
maintained between the pump and the upper Y-site.
1.
Close the in-line clamp on the IV infusion set. Press the door
latch to open the door.
2.
Release the flow stop mechanism by pushing the lever arm up
and to the right.
3.
Insert the IV infusion set from left to right into the slot provided,
following the flow direction label, avoiding any slack.
4.
Re-engage the flow stop mechanism by pushing the lever left
and down.
5.
Close the door. Use the recesses in the door to press firmly to
ensure the latch is correctly applied. Open in-line clamp on the
IV infusion set.
6.
Observe fluid chamber and check for no fluid flow.
8/77
1000SM00006 Issue 4