Configured options record sheet Alaris® GW Volumetric Pump
Enter the pump- specific information for your records on a copy of this page.
Description
Enable Volume / Time Infusions
Maximum Priming Volume
Clear Infusion Parameters
Maximum VTBI in MICRO Mode
Bolus Rate
Maximum Bolus Volume
Keep Vein Open Rate
Single Bubble Alarm Volume
Enable Secondary Infusions
Default Occlusion Pressure
Alarm Volume Level
Enable MICRO Mode
Maximum Infusion Rate
ASCII Mode for Comms
Odd Parity for Comms
Pump Address for Comms
Flow Sensor Connection Mode
Set-up of Current Time and Date
Language Selection
IrDA Communications Selection
Nurse Call Activation
Drops per ml of Fluid
Silent Mode
User select mode options
• Pressure limit Enabled
• Alarm volume Enabled
• Timed infusions Enabled
• Micro infusions Enabled
Flow sensor sensitivity level
*EnGL - English, FrAn - French, dEut - German, ItAL - Italian, ESPA - Spanish, SE - Swedish, nEd - Dutch.
Serial Number
Configured by
Approved by
Alaris® GW Volumetric Pump
Configuration Records
Configuration Records
Range
(ON / OFF)
(OFF, 1 - 40ml)
(ON / OFF)
(0.1ml - 999ml)
(1 - 999ml/h)
(OFF, 1 - 99ml)
(OFF, 1.0 - 5.0ml/h)
(50µL, 100µL, 250µL, 500µL)
(ON / OFF)
(Lo (250mmHg), Nor (350mmHg),
Hi (500mmHg)
(1 -7)
(ON / OFF)
(1 - 999ml/h)
(ON / OFF)
(ON / OFF)
(1 to 250)
(ON / AUTO)
(00:00 - 23:59)
(01/01/00 -31/12/99)
(EnGL, dEut, FrAn,ItAL, ESPA, nEd, SE)*
(ON / OFF)
(ON / OFF)
(1 to 200)
(ON / OFF)
(ON / OFF)
(ON / OFF)
(ON / OFF)
(ON / OFF)
(Nor, Hi)
73/77
Default
Setting
OFF
40ml
OFF
999ml
400ml/h
5ml
5.0ml/h
100µL
OFF
Hi
4
OFF
999ml/h
OFF
OFF
1
AUTO
N/A
EnGL
ON
ON
20
OFF
OFF
OFF
OFF
OFF
Nor
Software Version
Date
Date
1000SM00006 Issue 4