Door assembly

1.
Remove the air sensor assembly from the back of the door and retain the small screws for re-assembly later.
2.
Remove circlips and push out the two shafts that form the hinge of the door so that they clear the first part of the hinge.
Do not free the main pressure plate that is sprung on the door.
3.
The old door assembly will now come free from the front of the pump.
4.
Reassemble in reverse order.
When fitting the door, take care to keep the air sensor on the correct side of the door
and do not crush its wires.
(B) Door assembly
Item
Description
A
MAGNET DOOR
B
ASENA GW, KIT, DOOR
C
ASENA GW, KIT, FIXINGS(SCREWS,WASHERS,ETC)
Alaris® GW Volumetric Pump
Spare Parts Replacement Procedures
(A) Door magnet
3/77
(C) Circlips (x2)
(C) Screws (x2)
Air sensor
Part Number
1000ME01151
1000SP00253
1000SP00489
1000SM00006 Issue 4