Cardinal Health Alaris GW Technical & Service Manual page 53

Volumetric pump
Table of Contents

Advertisement

Electrical
Class I Equipment
Type CF Equipment (degree of protection
against electrical shock)
Electrical Safety
Electromagnetic compatibility (EMC)
Electrical safety
Dielectric strength
AC power supply
Battery power supply
Battery life
Battery charging
Environmental
Operating limits
Temperature
Humidity
Atmospheric pressure
Transport limits
Temperature
Humidity
Atmospheric pressure
IPX rating
Infusion
Priming the IV infusion set / Prime
Parameter
Priming rate
Priming volume
Starting the infusion / set-up
Infusion parameter
Flow rate
VTBI
VI
Administering a Bolus
Parameter
Bolus rate
Bolus volume
Max bolus volume after
release of hard occlusion.
*Non condensing.
**Measured in 0.1ml increments.
***Measured in 1ml increments.
Alaris® GW Volumetric Pump
Specifications
Specifications
Complies with EN60601-1-1 (IEC601-1) and EN60601-2-24 (IEC601-2-24).
Complies with EN60601-1-2 and EN60601-2-24.
EN60601-1 (IEC 601-1) - typical earth leakage current 40µA.
Proof strength test 1.5kV dc (live and neutral to earth).
Performance strength test 500V dc (live and neutral to earth).
115 VAC, 50/60Hz, 10VA (nominal).
230 VAC, 50/60Hz, 10VA (nominal).
Rechargeable NiMH (Nickel Metal Hydride). Automatically charges when the
Alaris® GW Volumetric Pump is connected to AC power.
>6 hours @ 25 ml/h, >4 hours @ 125 ml/h, >2 hours @ 999 ml/h with new, fully
charged cells with no loss of infusion accuracy; measured with 95% percentage
population / 95% confidence interval.
95% charge - < 24 hours (all conditions).
Min
+15°C
20%
700mbar
Min
-20°C
10%
500mbar
IPX 1
Range
Fixed; >999ml/h
0 - 40ml***
Micro
1.0ml/h - 99.9ml/h**
0.1ml - 99.9ml**
100ml - 999ml***
0.0ml - 99.9ml**
100ml - 9999ml***
Range
1 - 999ml/h***
0 - 99ml***
<0.6ml
3/77
Max
+38°C
90%*
1060mbar
Max
+50°C
95%*
1060mbar
Standard
1ml/h - 999ml/h***
1ml - 9999ml***
0ml - 9999ml***
1000SM00006 Issue 4

Advertisement

Table of Contents
loading

Table of Contents