Air sensor assemblies

1. Unplug the four-way connector for the two air sensors.
2. Unscrew the door air sensor from the back of the door and pull out the two wires through the hole to free the sensor.
(A) Door air sensor
3. Carefully push out the front panel air sensor, while moving the three retaining lugs towards the centre of the sensor, by
pressing gently on the encapsulated area with a screwdriver. Again, pull the two wires free from the front case.
4. Reassemble in reverse order.
4-way connector
(A) Door air sensor
The two air sensors are similar, but can easily be distinguished. The door sensor has a flange
with two countersunk holes in it. The front panel sensor has three sprung clips to hold it in
the case.
Insert the crimps into the four-way connector provided, as indicated by the following
diagram:
Item
Description
A
ASENA GW, KIT, AIR SENSORS
Alaris® GW Volumetric Pump
Spare Parts Replacement Procedures
Pin 1 Door Sensor White
Pin 2 Door Sensor White
Pin 3 Front Panel Sensor Blue
Pin 4 Front Panel Sensor Blue
1

3
4
37/77
(A) Front panel air sensor
Rear of front panel air sensor mounting
(A) Front panel air sensor
Part Number
1000SP00265
1000SM00006 Issue 4