Data transfer (continued)
Teach Learn (Software Versions VR1F and above)
1.
For both the teach and learn pumps in Technician Mode enable IrDA communications (Code 4), and ensure that ASCII / Binary
mode and parity bit options (Codes 38 and 3 respectively) are the same.
2.
Turn the teaching pump on in normal operation. Note: For multiple teach-learn operations, to avoid call-back alarm every
2 minutes, turn the teaching pump on in Technician Mode.
3.
Enter Technician Code 67 on the learning pump.
4.
Align the two IrDA ports on the pumps (optimum distance 50 mm).
5.
Depress the
key to initiate learning.
6.
A progress bar will travel across the learn pump.
7.
When successful, the learn pump will display 'PASS' .
8.
If the learning pump is unable to learn all configuration parameters then the display will show 'ConF' followed by a list of the
configuration parameters that could not be learnt; these will instead contain the factory default settings. This could occur if, for
example, the software version of the learning pump is newer than that of the teaching pump.
Possible Reasons for failure
Possible Reasons for failure
Event Log Download
A PC application known as the Event Log Download Utility (ELDU) is available to download the event log from the Alaris® GW
Volumetric Pump.
ELDU Operation
1.
Click on ELDU icon on PC.
2.
Click Accept to agree with Restrictions of Use and continue;
3.
Select Configure from drop-down menu;
4.
Select Setup Pump and choose Alaris® GW as pump type;
5.
Select Settings to select log to be downloaded;
6.
Check communications are set up as follows:
Required PC Comm port selected
Character type and parity match pump configuration
7.
Click OK to confirm
8.
Align the IrDA converter with the IrDA window (optimum distance 50 mm), or connect an RS 232 cable.
9.
Power up the pump by pressing the
10.
Click Download log from the main PC screen.
11.
Press Close when finished.
12.
Select File from drop-down menu and save file. Log may be printed here as required.
Alaris® GW Volumetric Pump
Routine Maintenance
Routine Maintenance
IrDA not enabled on both pumps;
IrDA not enabled on both pumps;
ASCII / Binary and parity bit options are not the same;
ASCII / Binary and parity bit options are not the same;
If the software versions are not compatible;
If the software versions are not compatible;
If the pump models are different;
If the pump models are different;
The line of sight between the IrDA windows was obstructed during data
The line of sight between the IrDA windows was obstructed during data
transfer.
transfer.
key.
/77
1000SM00006 Issue 4