Cardinal Health Alaris GW Technical & Service Manual page 57

Volumetric pump
Table of Contents

Advertisement

Electromagnetic Compatibility (continued)
The Alaris® GW Volumetric Pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the Alaris® GW Volumetric Pump should assure that it is used in such an environment.
Immunity Test
EN 61000-4- Electro-Static
Discharge (ESD)
EN 61000-4-4
Electrical Fast Transient, Burst
(EFT) (Note 3)
EN 61000-4-
Power Line Surge
(Note 3)
EN 61000-4-8 Power
Frequency Magnetic Field
(50/60 Hz)
EN 61000-4-11
Voltage Dips, Short
Interruptions, and Voltage
Variations
(Note 3)
Note 1—U
is the AC mains voltage prior to application of the test level.
T
Note 2—Compliance levels raised by EN 60601-2-24.
Note 3—Performed at the Minimum and Maximum Rated Input Voltage.
Note 4—Cardinal Health recommends using signal cables of less than 3 meters in length and this requirement is applicable only if signal cables are 3
meters or more in length. (EN 60601-1-2:2002, Clause 36.202.4)
Alaris® GW Volumetric Pump
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
EN 60601-1-
Test Level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV Line(s) to
Line(s)
±2 kV Line(s) to
Earth
3 A/m
<5 % UT (Note 1)
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Specifications
Compliance Level
±8 kV contact
Floors should be wood, concrete, or ceramic tile.
(Note 2)
If floors are covered with synthetic material, the
±15 kV air (Note 2)
relative humidity should be at least 30 %.
If connector testing exemption is used, the following
symbol for ESD sensitivity appears adjacent to each
connector. "Caution – Do Not Touch".
±2 kV for power
Mains power quality should be that of a typical
supply lines
commercial or hospital environment.
N/A (Note 4)
Mains power quality should be that of a typical
±1 kV Line(s) to
commercial or hospital environment.
Line(s)
±2 kV Line(s) to Earth
400 A/m 50 Hz
Power frequency magnetic fields should be at
(Note 2)
levels characteristic of a typical location in a typical
commercial or hospital environment.
<5 % UT
Mains power quality should be that of a typical
(>95 % dip in UT)
commercial or hospital environment.
for 0.5 cycle
If the user of the pump requires continued
operation during power mains interruptions, it is
recommended that the pump be powered from an
40 % UT
uninterruptible power supply or a battery.
(60 % dip in UT)
for 5 cycles
The pump does employ an internal short duration
battery.
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
7/77
Electromagnetic Environment – Guidance
1000SM00006 Issue 4