This section outlines the procedures for calibration of the Alaris® GW Volumetric Pump.
All of these calibrations should only be carried out by qualified biomedical engineers. If in any doubt about how to perform the tests,
in particular the pressure sensor calibration checks, contact your local Cardinal Health Service Centre who will be able to assist.
Calibration procedures
Recommended Calibration Equipment
Specialised test equipment is not required for the majority of the functional tests to be carried out on the pump. In order to calibrate
or verify the occlusion alarm point or volumetric accuracy, the following equipment will be necessary:
IV infusion set, suitable for the Alaris® GW Volumetric Pump with standard LUER lock taps for connecting to other test equipment.
If the standard sets are not available, it is possible to order a basic "test set" that is available from your local ALARIS® Service Centre
- part number 0000TG00074. Note that all sets should only be used for a single calibration operation.
Pressure gauge for measuring liquid pressure, with a full scale of 0-2 bar (0-1500mmHg) ±20mmHg.
either:
Class A 50ml glass burette with graduations down to 0.1ml and calibrated down to ± 0.05ml.
or:
Calibrated scales accurate to at least ± 0.01g.
Volumetric Accuracy Calibration (CODE 18)
Volumetric Accuracy Calibration (CODE 18)
The volumetric accuracy calibration routine is used for manufacture of the pump only.
The volumetric accuracy calibration routine is used for manufacture of the pump only.
Displaying the Volumetric and Pressure Calibration Values (CODE )
Use of this access code simply displays the calibration values stored in the software.
1.
Enter the access code 2.
2.
Press
to step through all Cal values.
3.
Switch the pump "OFF" if there are no further tests to be done at that time.
Battery Calibration (CODE 4)
The pump must be connected to the AC power source throughout the duration of the test. The test can be aborted at any time by
switching the pump off using the
pump. This calibration should only be performed on a fully charged battery. Ensure the pump is plugged into the mains for at least
4 hours before starting this procedure.
1.
Plug a mains lead into the pump.
2.
Prepare a fluid-filled looped tubing, load it into the pump and close the door.
Enter the access code 4.
3.
4.
The test will start automatically. The pump infuses at 999 ml/h internally switching to the battery power source to discharge the
battery; the air-in-line alarm is disabled. Throughout this test the time indicator will increase in minute intervals.
5.
As soon as the pump detects that the battery is discharged, the time shown on the main display will stop increasing and begin to
flash.
6.
If the time is greater than two hours and the low point voltage value is within the allowable range of 6.5 to 7.8 volts, then the
display shows "PASS" / "xx:xx" / "bx.x", the pass indication, elapsed time and battery low point value in volts. Otherwise the
display shows "FAIL" / "xx:xx" / "bx.x".
7.
Press the
key. When the
If the pump is switched off before the
If the pump is switched off before the
test will have to be repeated. When
test will have to be repeated. When
entry mode and flash CodE, followed by 0, this allows you to begin other tests. Switch the pump
entry mode and flash CodE, followed by 0, this allows you to begin other tests. Switch the pump
OFF if there are no further tests to be done at that time.
OFF if there are no further tests to be done at that time.
8.
If the pump fails the calibration, try to charge the battery, and repeat the test. If this fails again then replace the battery and / or
Power Supply Unit.
Alaris® GW Volumetric Pump
Configuration & Calibration
Configuration & Calibration
key; no change is made to the battery low point calibration value stored previously in the
key is pressed, the low point calibration value will be stored.
13/77
key is pressed, the calibration value will be lost and the
key is pressed, the calibration value will be lost and the
is pressed, the pump will revert to the technical service
is pressed, the pump will revert to the technical service