Cardinal Health Alaris GW Technical & Service Manual page 12

Volumetric pump
Table of Contents

Advertisement

Configurable options (continued)
Code
Description
39
Odd Parity Comms
40
Pump Address Comms
41
Flow Sensor Connection Mode
42
Set-up of Current Time and Date
44
Language Selection
45
IrDA Communications Selection
46
Nurse Call Activation
47
Drops per ml of Fluid
48
Silent Mode
User Select Mode Configuration:
• Pressure Limit Enabled
49
• Alarm Volume Enabled
• Timed Infusions Enabled
• Micro Infusions Enabled
50
Flow Sensor Sensitivity Level
Reset all configurable options to
200
default
201
Reset EEPROM data
202
Repair EEPROM data
*Single Bubble Alarm Volume
Although an individual bubble may not exceed the pre-programmed threshold, the
accumulative volume of bubbles, in a 1 minute window, may be sufficient to initiate an air-in-
line alarm, indicated by an "Air OCCL" message.
Alaris® GW Volumetric Pump
Configuration & Calibration
Configuration & Calibration
Default
OFF
1
AUTO
EnGL
ON
ON
20
OFF
OFF
nor
-
-
-
Before making any amendments to configuration settings:
Care should be taken to document existing configuration settings to enable changes
to be reverted if required. Configuration requirements may vary from ward to ward
therefore care should be taken to ensure any configuration settings are appropriate
for the ward concerned and users are aware of any changes to configuration settings
prior to use.
Subsequently, sharing of Alaris® GW Volumetric Pumps between wards may be
inappropriate.
OFF:
Disable odd communications parity bit generation.
ON:
Enable odd communications parity bit generation.
Set pump address used for communications (1 to 250).
AUTO: Pump automatically detects flow sensor if connected.
ON:
Pump will only operate with a flow sensor connected.
Set current time (00:00 to 23:59), and date (01/01/00-31/12/99) for
event logging. Does not automatically adjust for Summer time.
Set language used (EnGL, dEut, FrAn, ItAL, ESPA, nEd, SE).
OFF:
RS232 comms enabled.
ON:
IrDA comms enabled.
OFF:
Disable activation state of the nurse call (active low output from
pump).
ON:
Enable activation state of the nurse call (active high output
from pump).
Select number of drops per ml of fluid (1 to 200). Defined by the type of
set. Reference the packaging of IV infusion set.
OFF:
Audible response to a key press is given.
ON:
No audible response to a key press is given.
OFF:
Disable mode.
ON:
Enable mode.
nor:
Normal sensitivity.
Hi:
High sensitivity.
Resets all configurable options to factory default.
Reset EEPROM data code to a defined state if EEPROM checksum error.
NOTE: The pump will need to be returned to factory for reconfiguration
if this option is used.
Detects and repairs any corrupted memory segments, resetting any
repaired areas back to the factory defaults.
NOTE: The pump will need to be returned to factory for reconfiguration
if this option is used.
1/77