Psu & Comms Pcb - Cardinal Health Alaris GW Technical & Service Manual

Volumetric pump
Table of Contents

Advertisement

PSU & Comms PCB
1.
Unplug the mains inlet, the battery and the 4-way cable assembly from the PSU and Comms. PCB and 8-way connector and flow
sensor connector.
2.
Remove the four securing screws that hold in the PCB and remove the PCB from the rear case.
3.
Reassemble in reverse order.
(D) Screws (x4)
(C) 4-way cable assembly
Flow sensor and speaker cables have been removed for clarity.
Power supply kit (1000SP0047) includes new RS3 connector and battery
clamp, as previous parts are not compatible with new PSU. Replace RS3
connector and battery clamp if required.
Item
Description
A
ASENA GW, KIT, POWER SUPPLY UNIT (PSU)
B
ASENA GW, ASSY, SERIAL COMMS CABLE
C
ASSY CABLE 4 WAY
D
ASENA GW, KIT, FIXINGS(SCREWS,WASHERS,ETC)
Alaris® GW Volumetric Pump
Spare Parts Replacement Procedures
43/77
(B) Serial comms cable
(A) PSU
Part Number
1000SP00427
1000SP01135
1000SP01076
1000SP00489
1000SM00006 Issue 4

Advertisement

Table of Contents
loading

Table of Contents