Mains Inlet - Cardinal Health Alaris GW Technical & Service Manual

Volumetric pump
Table of Contents

Advertisement

Mains inlet

(D)
Magnet (item D) can be held in place with a plastic
covered proximity magnet placed over IR window.
Magnet will be held in place with mains inlet retainer
(item B).
Item
Description
A
ASENA GW, ASSY, MAINS INLET
B
ASENA GW, ASSY, MAINS INLET RETAINER
C
GASKET MAINS INLET
D
MAGNET IR DETECT
E
ASENA GW, KIT, FIXINGS(SCREWS,WASHERS,ETC)
Alaris® GW Volumetric Pump
Spare Parts Replacement Procedures
(A) Mains inlet assembly
47/77
1.
Remove 2 screws securing Mains inlet and retainer.
2.
Remove Mains inlet and retainer.
3.
Remove magnet.
4.
Reassemble in reverse order.
(C)
Part Number
1000SP01134
1000ME01443
1000ME01299
1000ME01303
1000SP00489
(E)
(B)
1000SM00006 Issue 4