Pressure Sensor Assembly - Cardinal Health Alaris GW Technical & Service Manual

Volumetric pump
Table of Contents

Advertisement

Pressure sensor assembly

1.
Remove the three screws that hold the pressure sensor assembly in place. Unplug the assembly from the Control PCB. Carefully
remove the pressure sensor from the case.
2.
Reassemble in reverse order. Tighten the larger 2 screws first, then the smaller screw.
(D) Screws (x3)
(C) Backstop
Item
Description
A
ASENA GW, KIT, PRESSURE SENSOR 230V
B
ASENA GW, KIT, PRESSURE SENSOR 110V
C
ASENA GW, ASSY, BACKSTOP/MEMBRANE CLAMP
D
ASENA GW, KIT, FIXINGS(SCREWS,WASHERS,ETC)
Alaris® GW Volumetric Pump
Spare Parts Replacement Procedures
Side View
38/77
Underside View
(A or B) Pressure sensor and encoder assembly
Part Number
1000SP00256
1000SP00330
1000ME01592
1000SP00489
1000SM00006 Issue 4