Test procedures (continued)
Volumetric Accuracy Verification Test (CODE )
Important Notes:
Important Notes:
The balances need to be switched on for 30 minutes prior to use to enable the electronics to
The balances need to be switched on for 30 minutes prior to use to enable the electronics to
warm up and settle.
warm up and settle.
Always use new test tubing for each test. If the test ever has to be repeated, a new set of tubing
Always use new test tubing for each test. If the test ever has to be repeated, a new set of tubing
must be used.
must be used.
Do not move the desk during testing, it will upset the balance readings.
Do not move the desk during testing, it will upset the balance readings.
This test is used to confirm that the pumping accuracy of the system as a whole, including the tubing, is within the specified limits.
In the most controlled conditions, a needle should be used to pump liquid into the weighing beaker to prevent liquid touching the
sides of the beaker and to provide some back-pressure so that leaks/overflows do not affect the readings. As a result of these and
other errors, if the system fails just marginally, it is worth performing the test a second time. If it still fails, return the pump to your local
ALARIS® Service Centre for further analysis. The head height on IV infusion set should be approximately 30cm.
Setup A
Setup A
30cm
30cm
1.
Enter access code 5. The pump will initially display "
2.
If it is necessary to prime the set, press and hold the
ignoring any air-in-line alarms.
3.
Release
once the set is primed and the pump will again display "
Alaris® GW Volumetric Pump
Routine Maintenance
Routine Maintenance
or...
or...
Class A Burette (50ml)
Class A Burette (50ml)
0.1ml graduations
0.1ml graduations
Burette clamp
Burette clamp
30cm
30cm
Equipment stand
Equipment stand
Table or bench
Table or bench
".
button. The pump will display "
18/77
Setup B
Setup B
" and allow the set to be primed,
".
Needle
Needle
Beaker
Beaker
Scales (±0.01g)
Scales (±0.01g)
1000SM00006 Issue 4