Figure 5 - Flow Pressure Nomogram - Cardinal Health Infant Flow SiPAP Operator's Manual

Table of Contents

Advertisement

Operator's Manual
Flow / Pressure Relationship
The Infant Flow
enriched gas flow and airway pressure. A nomogram illustrating the relationship
between constant airway pressure and flow settings is shown in Figure 5. For
example, 8 L/min gas flow provides approximately 5 cmH
Note
Individual devices have a tolerance of up to ± 10% from that illustrated in the
nomogram and in particular, at pressures below 2 cmH
675-101-101
Revision H
®
SiPAP™ is subject to a direct relationship between the controlled
Figure 5 – Flow Pressure Nomogram
O.
2
O.
2
15

Advertisement