Table 2
For all ME DEVICES and ME SYSTEMS Guidance and Manufacturer's Declaration – Electromagnetic immunity
The Veroval
®
wrist blood pressure monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the
Veroval
wrist blood pressure monitor should assure that it is used in such an environment.
®
Electromagnetic interference
immunity tests
Electrostatic discharge (ESD) IEC
61000-4-2
Power frequency (50/60 Hz) mag-
netic field IEC 61000-4-8
NOTE: U
is the mains alternating voltage before use of the test levels.
T
IEC 60601 test level
± 6 kV contact ± 8 kV air
3 A/m
Compliance Level
± 6 kV contact ± 8 kV air
3 A/m
95
English
Electromagnetic Environment – Guidance
Floors should be wood, concrete or ceramic
tile. If the floor has a synthetic material
covering, the relative humidity should be at
least 30%.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.