Hartmann Veroval BPM25 Instructions For Use Manual page 96

Wrist monitor
Hide thumbs Also See for Veroval BPM25:
Table of Contents

Advertisement

Available languages

Available languages

English
GB
Table 3
For all NON-LIFE-SUPPORT medical-electrical (ME) DEVICES and ME SYSTEMS.
Guidance and Manufacturer's Declaration – Electromagnetic immunity
The Veroval
wrist blood pressure monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the
®
Veroval
wrist blood pressure monitor should assure that it is used in such an environment.
®
Electromagnetic interfer-
ence immunity tests
Radiated RF disturbances
IEC 61000-4-3
IEC 60601 test level
Compliance Level
3 V/m 80 MHz to
3 V/m
2.5 GHz
Electromagnetic Environment – Guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the Veroval
pressure monitor, including its cables, than the recommended
separation distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2.5 GHz
Where P is the maximum output power of the transmitter in
watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey
pliance level in each frequency range
Interference may occur in the vicinity of
equipment marked with the following symbol:
96
wrist blood
®
a
, should be less than the com-
b
.