Table 4
For all NON-LIFE-SUPPORT medical-electrical (ME) DEVICES and ME SYSTEMS.
Recommended separation distances between portable and mobile RF communications equipment and the Veroval
sure monitor
The Veroval
wrist blood pressure monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
®
The customer or user of the Veroval
between portable and mobile RF communications equipment (transmitters) and the Veroval
according to the maximum output power of the communications equipment.
Rated maximum output power of trans-
mitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
wrist blood pressure monitor can help prevent electromagnetic interference by maintaining a minimum distance
®
Separation distance according to transmitter frequency (m)
80 MHz to 800 MHz
d = 1.2
0.12
0.38
1.2
3.8
12
wrist blood pressure monitor, as recommended below,
®
800 MHz to 2.5 GHz
d = 2.3
0.23
0.73
2.3
7.3
23
98
®
wrist blood pres-