VWR avantor CHEMI Instruction Manual page 27

Table of Contents

Advertisement

If you would like the software to pick a generic chemiluminescent reagent as default
please check the box. Alternatively, you can select the chemiluminescent reagent you
wish to use as default by using the drop down menu.
Select to 'Allow binning by Default' use the drop down menu to select binning level.
Binning will increase sensitivity but reduce resolution.
Enter default values for exposure time. Check the box to use the same exposure time
for all images
Select the number of images you wish to capture and whether you would like to add
the contents of previous image. This will add the each captured image to the previous
image. This will add each captured image to the previous image improving the signal to
noise ratio.
Signal Accumulation Calculator settings
Check box to image a visible or colour marker for Signal Accumulation Calculator.
If you would like the software to pick a generic chemiluminescent reagent as default
please check the box. Alternatively, you can select the chemiluminescent reagent you
wish to use as default by using the drop down menu.
Select to 'Allow binning by Default' use the drop down menu to select binning level.
Binning will increase sensitivity but reduce resolution
®
26 vwr.com