Maritime G6100 Operation Manual page 94

Water monitoring system
Table of Contents

Advertisement

Appendix
MI6100-0004
System Type
Water Monitoring System
Task:
Turbidity Wiper Unit Replacement
Task
This instruction describes the replacement of the Turbidity Wiper Unit. The
wiper unit replacement must be carried out by a trained technician.
description:
Difficulty 1-5
Overhaul Interval
When necessary
Spare Parts
Part no:
Description:
02805
Turbidity Wiper Unit
Instructions
Steps
Make sure the power supply is cut off prior
to replacing the wiper unit.
This is done by switching off the main
automatic fuse found inside the control
cabinet.
Remember to install the wiper unit nut
prior to installing the cable.
Notes:
Estimated Time Consumption
15 min
Tools
Qty.
Item:
Small size flat headed screwdriver
Medium size flat headed screwdriver
2 x 8mm spanners
Photos:
Wiper unit electrical connections
MI6100-0004
Page 94 of 110