Water Monitoring System
6.3.4 Validation of G6110 PAH Module
Figure 6-3: G6110 PAH sensor, measuring chamber & calibration chamber & bushing &
bushing tool
1. Remove the PAH chamber bushing from the PAH measuring chamber using the PAH
chamber bushing tool – Figure 6-4. Clean the bushing thoroughly.
Figure 6-4: PAH Chamber Bushing Placement
2. Press the PAH area on the home screen. Press Calibration Mode – Login – press
Calibration Mode again and wait until the status changes to "Calibration Mode
Enabled".
Based on the sensor range, please select the validation liquid according to Table 6-1.
3. Take the syringe and attach the needle. Break the neck of the ampoule – fill the syringe
with the entire content.
Ver.5.0 – Revision March 2022
Page 43 of 110