Specifications
Accuracy
pH/Temperature Module – Type G6130
Legislation
Approvals
Measurement Technology
Measurement Range
Power Supply
Wetted Materials
Signals
Accuracy
Sampling Systems (Optional)
Pressure Reduction Cabinet – WMR
Equipment
Connections
Electrical Connections
Sample Temperature
Sample Purity
Sample Condition
Dimensions & Weight
Wetted Materials
Sample Pressure
Enclosure Material
Pump Cabinet – WMP
Equipment
Max ± 2 NTU
Compliance towards MEPC. 259(68) has been
demonstrated under the surveillance of various class
societies
Refer to Green Instruments website
According to BS EN ISO 60746-1 &
BS EN ISO 60746-2:2003
0...14 pH units; 0...50°C
Refer to Electrical Drawings
AISI 316 or PPS, glass, PVDF & PVC
2 Analog 4...20 mA (Passive)
Max ±0.2 pH unit
20 mesh strainer, Pressure Reducing Valve, Relief
Valve
ON/OFF Valve, Flow Regulating Valve, De-bubbler, &
Manometer
Refer to the PI diagram & the Installation Layout
Refer to the Electrical Drawings
0...50
o
C
Particle size of max. 2 mm
Avoid air and particles in the sampling system
Refer to the installation layout
AISI 316, Acetal, PTFE, PVDF, PC
2...10 bar
AISI 304 Painted or Polycarbonate
20 mesh strainer, Flow Regulating Valve, Pump
Debubbler, and Manometer