Maritime G6100 Operation Manual page 64

Water monitoring system
Table of Contents

Advertisement

Troubleshooting
Trouble
No display despite
power supply and circuit
breaker are OK
Incorrect indication of
pH value
Incorrect indication of
Turbidity value
Incorrect indication of
PAH value
Fluctuations of readings → Possibly caused by air bubbles. Check the sampling line. Check
No data log
No historical trend curve
Increased pump
temperature
Possible Cause →Action
→The display is defect →Reboot, repair, or exchange the display.
→ Try to clean the sensor and chamber manually.
→ Try to validate the pH Sensor.
→ Change the pH Electrode.
→ Try to clean the sensor and chamber manually.
→ Try to validate the Turbidity Sensor.
→ Water coloration might restrict light passing through the sample.
Try to minimize the water coloration.
→ Check the wiring from the Turbidity sensor to the Turbidity
Analyzer.
G6110: Check/clean the sensor, bushing, and chamber thoroughly.
G6111: Check/clean the sensor and chamber thoroughly.
→ Check expiration date of PAH sensor certificate
→ Validate the PAH Sensor
→ Check/remove bubbles in calibration liquid.
→ Check calibration liquid expiration date and temperature
→ Do not expose PAH calibration liquid to UV light for more than a
few minutes since UV light degrades PAH molecules
that the de-bubbler is functioning.
→ Wait for the trend to generate the curves.
→ Check if the SD card is in position.
→ If yes, check the SD card's lock mode.
→ Replace the SD card if it is defect.
→ Pump/motor rotation is hard. Maybe water in bearing or foreign
material block pump → inspect motor and pump to identify reason
for blockage or other factors that stops the motor or pump from
rotating freely.
Page 64 of 110