Menu Structure
SIGNAL SETUP
There are 5 signal setup pages: PAH, pH, Turbidity, Temperature and Flow. Each page
allows the user to edit the settings for each signal.
G6110/G6111 PAH Setting Example
Please find the details of the PAH signal setup page below.

Figure 4-13: Signal Setup Page - Low Range PAH Setting

Low Reference: The Low ampoule used for PAH calibration.
High Reference: The High ampoule used for PAH calibration.
Low Raw: The RAW signal captured when performing the zero calibration.
High Raw: The RAW signal captured when performing the span calibration.
Refresh: Stores edited values.
Please note the system is calibrated at the factory. The factory calibration values can be
found in the Test and Configuration Sheet included with each system. To recreate the
factory settings, please use the values on the Test and Configuration Sheet. This can be
done by entering the relevant value on the signal setup page.
Flow Setting Example

Figure 4-14: Signal Setup Page - Flow Setting

Low Reference: A fixed value determined by the range of flow sensor.
High reference: A fixed value determined by the range of flow sensor.
Page 24 of 110