Table 6-1: Validation/Calibration Liquid Selection; Table 6-2: Validation Tolerance Table - Maritime G6100 Operation Manual

Water monitoring system
Table of Contents

Advertisement

Validation & Calibration
ATTENTION
Cleaning of the G6110 calibration chamber, PAH chamber bushing and the optic(s)
It is crucial that the calibration chamber, PAH chamber bushing for G6110 and the optic(s)
is/are clean prior to validation. Any contaminants present in the chamber, PAH chamber
bushing or on the lens will result in a faulty validation/calibration.
It is vital to wipe the sensor dry prior to and between any validation & calibration attempt.
VALIDATION/CALIBRATION LIQUID SELECTION
Module Type
Calibration
G6110
Low
Range
G6111
G6110
High
Range
G6111
G6110
Dual
Range
G6111

Table 6-1: Validation/Calibration Liquid Selection

Standard Concentration
50
100
200
400
800

Table 6-2: Validation Tolerance Table

ATTENTION
Do not leave the sensor in the PAH calibration liquid more than 15 seconds since UV light
degrades the PAH calibration liquid.
Calibration Liquid µg/l
Low Range
Calibration
Low
High
Calibration
Calibration
0
100
-
-
0
100
Validation Tolerance µg/l
Lower Tolerance
High Range Calibration
Low
Middle
Calibration
Calibration
-
-
0
400
100
400
-
45
95
160
360
760
Validation
Liquid µg/l
Low
High
Range
Range
High
-
50
-
800
-
400
800
50
200
Upper Tolerance
55
105
240
440
840
Page 42 of 110