Maritime G6100 Operation Manual page 11

Water monitoring system
Table of Contents

Advertisement

Water Monitoring System
Measurement Technology
Measurement Range
Power Supply
Wetted Materials
Signal
Accuracy
Turbidity Module – Type G6120
Legislation
Approvals
Measurement Technology
Measurement range
Power Supply
Wetted Materials
Signal
Ver.5.0 – Revision March 2022
The technique used is UV induced fluorescence with the
capability of detecting the PAH concentration in μg/l
(ppb).
Excitation wavelength: 254 +/-10nm
Detection wavelength: 360 +/- 50nm
The PAH measurement is compensated for the effects of
turbidity by means of a turbidity compensation formula
and input from the turbidity sensor module G6120.
Low range: 0...100 µg/l phenanthrene equivalence, or
High range: 0...800 µg/l phenanthrene equivalence or
Dual range: 0...100/800 µg/l phenanthrene equivalence
To be specified upon order.
Refer to Electrical Drawings
G6110: AISI 316, acetal, Titanium gr.2, sapphire glass,
PVDF and NBR
G6111: PVC, fused silica glass, AISI 316, PVDF and
NBR
G6110: RS-232 and Analog 4...20 mA (Active)
G6111: RS-232 and Analog 0/4...20 mA (Active)
Max ±5% of sensor range or
Max ±5% of the nominal standard test concentration,
which value is not less than 80% of the sensor range.
Compliance
towards
demonstrated under the surveillance of various class
societies.
Refer to Green Instruments website
According to ISO 7027
0...400 NTU
Refer to Electrical Drawings
AISI 316, acetal, PVDF, NBR and glass
Analog 4...20 mA (Active)
MEPC.
259(68)
Page 11 of 110