Validation & Calibration
6.4.1

Storing a Calibration Point of G6110 PAH Module

Figure 6-6: G6110 PAH sensor, measuring chamber & calibration chamber & bushing &
bushing tool
1. Remove the PAH chamber bushing from the PAH measuring chamber using the PAH
chamber bushing tool – Figure 6-7. Clean the bushing thoroughly.

Figure 6-7: PAH Chamber Bushing Placement

2. Press the PAH area on the home screen. Press Calibration Mode – Login – press
Calibration Mode again and wait until the status changes to "Calibration Mode
Enabled".
Based on the sensor range, please select the calibration liquid according to Table 6-3
have been in use.
Page 46 of 110