Calibration Of The Pah Module; Table 6-3: Validation/Calibration Liquid Selection - Maritime G6100 Operation Manual

Water monitoring system
Table of Contents

Advertisement

Water Monitoring System
2. Take the syringe and attach the needle. Break the neck of the ampoule – fill the syringe
with the entire content.
3. Slowly inject the content into the PAH calibration chamber, taking care that no air is
mixed into the validation liquid. Look into the chamber to check and remove air
bubbles.
4. Place the cleaned PAH sensor in the calibration chamber in an upright position. Wait for
the display of PAH value to approach the value of the used ampoule. If the displayed
value is within the tolerance in Table 6-2, the validation is successful, and a calibration
is not required. If the validation is not successful, refer to Figure 6-2.
5. Empty out the validation liquid from the calibration chamber and wipe it dry with a new
cleaning pad. Clean the PAH sensor.

6.4 Calibration of the PAH Module

ATTENTION
Do not leave the sensor in the PAH calibration liquid more than 15 seconds since UV light
degrades the PAH calibration liquid.
Note
G6110: Normally all local calibrations are done using the PAH calibration chamber
with the cleaned bushing installed. However, the PAH measuring chamber can be
used if the PAH calibration chamber is missing.
Module Type
Calibration
G6110
Low
Range
G6111
G6110
High
Range
G6111
G6110
Dual
Range
G6111

Table 6-3: Validation/Calibration Liquid Selection

Ver.5.0 – Revision March 2022
Calibration Liquid µg/l
Low Range
Calibration
Low
High
Calibration
Calibration
0
100
-
-
0
100
High Range Calibration
Low
Middle
Calibration
Calibration
-
-
0
400
100
400
-
Validation
Liquid µg/l
Low
High
Range
Range
High
-
50
-
800
-
400
800
50
200
Page 45 of 110