Water Monitoring System
The following signs indicate an electrode reaching its end of lifetime. The electrode must
be replaced whenever one of the below signs are identified:
• The pH/Temperature sensor is difficult to calibrate. Storing a calibration point
should take max 180 seconds. The electrode must be replaced if longer time is
required.
• The pH/Temperature sensor was calibrated successfully, but the output drifts within
15 minutes after the calibration. To test drift please immerge the sensor in e.g. clean
tap water – note the initial displayed pH value – leave the sensor in the water for
15 minutes - then note the displayed value after 15 minutes.
• The pH/Temperature sensor gives a fixed output of around pH 7
• The pH/Temperature sensor gives no output.
• The glass tip of the electrode is damaged.
• Water has entered the electrical connection of the electrode.
• If the pH sensor flashes every second, the calibration is not accepted.
Changing of the Electrode is easily done by following the maintenance instruction
MI6100-0007.
Note
It is recommended to keep an electrode in stock as a spare part.
Ver.5.0 – Revision March 2022
Calculate drift:
Drift = "Initial value" ÷ "15-minute value"
Max drift allowed is +/-0.2 pH units
Page 71 of 110