Short instructions for use 6 Hygiene

6.11 Check the OXYGENAL 6 concentration

▶ Test the standard concentration 0.02% H
with 39 ml water and 1 ml treated water.
▶ Immerse a Merckoquant Peroxide Test Strip in the liquid.
O
(200ppm): Fill a measuring cup (40 ml)
2
2
40