Short instructions for use 8 Warning messages
Item
Safety shutdown actuated
no.
Bench
⑤
Kickplate
⑥
⑦
Seat base
Stirrup on foot control
⑧
▶ To deactivate an active safety shutoff, remove the trigger from the range of move-
ment of the chair.
8.2 Troubleshooting
Malfunction
Nothing works.
Nothing works.
Cause
Main switch is off.
The main fuse is
tripped.
Remedy
▶ Turn on main switch.
▶ Unplug the unit from the mains.
▶ Check and replace, if required, the main
service fuse.
The main service fuse is situated next
to the master switch.
▶ Use a screwdriver to open the bayonet
lock, and change the microfuse.
(220, 230, 240 V AC: T 6.3 H Mat. no.
0.223.2783);
(100, 110, 120, 130 V AC: T 10 H Mat.
no. 1.007.2529).
▶ The re-close the bayonet closure with
the screwdriver.
Display LED on the as-
sistant element
45