Short instructions for use 6 Hygiene
6.10 Replace the OXYGENAL 6 bottle
Only KaVo OXYGENAL 6 (Mat. no. 0.489.3451) should be used for KaVo water germ re-
duction systems.
▶ Wipe up any spilled OXYGENAL immediately in order to prevent discolouration of the
surfaces.
▶ Remove the cover from the unit base.
▶ Unscrew the cover of the new OXYGENAL 6 bottle.
▶ Swing out the empty OXYGENAL 6 bottle from the unit base, and pull it down off the
adapter.
The touchscreen shows a status message when the bottle is taken out of the unit. In or-
der to display the "No Oxygenal bottle" status message, tap first the arrow and then the
status message. Then the required steps can be done.
▶ Press the new bottle onto the adapter, and swing the bottle into the unit base.
▶ Putting the cover back on the unit base.
Note
Completely empty packages can be disposed of at recycling facilities.
See also :Safety data sheet
39