Short instructions for use 6 Hygiene
▶ During the active phase, do not place hoses in holders and, if applicable, do not
switch off the treatment centre.
▶ Wait for the end of the active phase or switch on the treatment centre.
ð A progress bar indicates the cleaning steps:
▶ Wait for the cleaning programme to be completed.
ð The following message is displayed after completion of the cleaning programme:
▶ After completion of the "Weekly" cleaning programme, place hoses in holders.
6.7 Intensive germ reduction during prolonged periods of non-
use
During extended periods of non-use, e.g. semester breaks at university, "Weekly" clean-
ing must be run during the time of non-use. The germ reduction liquid remains inside the
system during the period of non-use.
Before the period of non-use
▶ Start the "Weekly" cleaning programme and switch it off during the active phase after
an acoustical signal within the dwell time of 30 minutes.
▶ To prevent damage due to hoses getting kinked, remove hoses from the hygiene
centre and place them in the dentist holder.
▶ Cover the treatment unit during extended periods of non-use.
After the period of non-use
▶ Turn on the central compressor.
35