Care instructions ESTETICA E30
3 Replenishing and checking the disinfectant | 3.2 Check the OXYGENAL 6 concentration
Note
Substances remaining in the dosing head are not contamination, but rather
agents that are deposited there during extended periods of non-use after
condensation of the water. These deposits redissolve after several dosing
processes.
▶ Screw the intensive germ reduction bottle tightly to the treatment centre
through a clockwise rotation.
ð
Apply compressed air to the intensive germ reduction bottle.

3.2 Check the OXYGENAL 6 concentration

▶ Fill the tumbler with water.
▶ Use Merckoquant peroxide test strips 100 to 1000 mg/l H
0.229.5003) to determine the OXYGENAL 6 concentration in the water
germ reduction system. The OXYGENAL 6 concentration should be in the
range from 150 to 250 mg/l H
O
in normal operation.
2
2