Additional Warning Messages - KaVo ESTETICA E70 Vision Short Instructions For Use

Hide thumbs Also See for ESTETICA E70 Vision:
Table of Contents

Advertisement

Short instructions for use 8 Warning messages
Malfunction
The treatment unit is
not connected to the
wireless foot control.
The treatment unit
emits a tone.
The wireless foot
control does not re-
turn to centre posi-
tion.

8.3 Additional warning messages

Malfunction
A signal is emitted
every 10 seconds
and a status mes-
sage is shown.
Ten beeps are is-
sued.
A melody sounds.
Error messages on the touchscreen
Malfunction
Status message
""Unit controller not
recognised""
Cause
The wireless foot
control is turned off.
The wireless foot
control is out of
range.
Malfunction or low
battery
Cause
The oxygenal con-
tainer is empty.
The Oxygenal con-
tainer is too full.
The amalgam separ-
ator CAS1 is almost
full (95%).
The CAS1 amalgam
separator is defect-
ive.
Cause
Cable or electronics
issue.
Remedy
▶ Check the on/off switch on the foot
control and turn it on if necessary.
▶ Move the wireless foot control into the
range of the treatment unit.
▶ Check the status display of the foot
control.
Yellow: low battery
No display: malfunction
▶ Charge battery.
▶ Charge battery.
Remedy
▶ Refill the Oxygenal container.
▶ Stop filling the Oxygenal container.
▶ Exchange the amalgam container.
▶ Also refer to: Instructions for use for
the CAS 1 or
▶ Call a Service technician.
Remedy
▶ Call a service technician.
47