Figure 12 Electrical Connections For 115 V - M&C PSS Series Instruction Manual

Gas conditioning unit
Hide thumbs Also See for PSS Series:
Table of Contents

Advertisement

Figure 12
Electrical connections for 115 V
Option "temperature controller for heated sample lines":
For the electrical supply and control of a heated sample line with PT-100 a 7-pin plug at the side of the unit is
available. The connection power is max. 6 A, 1380 W for the 230 V version or 6 A, 690 W for the 115 V-version.
The maximum length of the heated sample line which can be used is calculated as follows:
Max. connection power [W]
L [m] = ––––––––––––––––––––––––––––––––––––
Power consumption of heated sample line [W/m]
www.mc-techgroup.com
SS5C, SS5C/2, SS5C/3 | 1.00.00