INTRODUCTION

The unit, which is mounted on an aluminium plate, is a complete sample gas conditioning unit suitable for
continuous use and can be easily integrated into analytical systems. This minimizes the effort for the procurement
of individual components and small parts as well as the assembly of the individual parts.
The compact design only requires a small amount of space. The system is ready for operation in just a few minutes.
7
FUNCTION OF THE M&C JET-STREAM HEAT EXCHANGER
The ECPX000C gas coolers, specially developed for analysis technology, are designed for maximum flow rates of
up to 350 Nl/h. They are also installed as system assemblies in the SS5C, SS5C/2 and SS5C/3.
Jet-Stream heat exchangers are available in Duran glass, stainless steel (316Ti) and PVDF (polyvinylidene fluoride).
The selection of the appropriate heat exchanger material is customer specific. The heat exchangers are easily
accessible and easily replaceable in a heat-insulated cooling block. Figure 2 shows a schematic diagram of the
heat exchanger function.
Figure 2
Diagram of the heat exchanger function
www.mc-techgroup.com
Sample gas
out
+5°C
[41 °F]
SS5C, SS5C/2, SS5C/3 | 1.00.00
Sample gas
in
Cooling-
block
Condensate
out
9