PARAMETER SETTING

If you tap the
-key after entering the PIN (see chapter "11.2 PIN Entry"), the set point for the cooler temperature
appears first. If the
-key is pressed and held for a short moment, the display changes to code entry. Here you
can enter the codes that belong to the respective parameter settings.
You can also access the parameter setting from the main menu. To do this, keep the
code entry is displayed. The PIN must be active in this case.
Note
A code can have up to 3 digits. Single-digit codes are device-specific, two-digit codes are important basic settings
(tens digit corresponds to the channel number), three-digit codes are used for calibration (hundreds digit: "2"
stands for LA, "3" for mA calibration).
The PIN is valid for 15 minutes. If you exceed this time, the display field reappears with "0000" when you press and
hold the
-key. The PIN must be entered there again (see chapter "11.2 PIN Entry").
If you enter an invalid code and press the
20
The codes for parameter settings can be found in the appendix of this
instruction manual.
To adjust the device parameters, the code belonging to the parameter must
be entered and confirmed in this display. The left digit of the code entry is
blinking. The first digit can be entered here.
Use the
und
keys to set the individual digits.
-key, the display returns to the cooler temperature.
SS5C, SS5C/2, SS5C/3 | 1.00.00
-keys to switch between the digits and the
-key pressed until the