Table of Parameter Codes
No.
Description:
1
Software version
5
Brightness setting of
display
10
LA on/off switching
11
Sensitivity LA
12
HIGH dT
13
LOW dT
14
Hysteresis
(temperature alarm)
15
mA range selection
70
Operating time in
days
84
Fan speed
210
Calibration LA dry
211
Calibration LA wet
212
Calibration LA open
/ cable break
213
Display LA dry
214
Display LA wet
215
Display LA open
310
mA
output
Calibration
311
mA output HIGH
Calibration
www.mc-techgroup.com
Default:
Range:
5
0 - 9
0
0,1,2
2
1-7
3
2 - 8 [°C]
3
2 - 8 [°C]
2
1,2 [°C]
4-20
0-20 / 4-
20 [mA]
1
0-5
LOW
1 mA
20 mA
SS5C, SS5C/2, SS5C/3 | 1.00.00
Note:
Brightest display setting is 9
0=off; 1= without cable break detection; 2= with cable
break detection
The higher the value is, the sooner the alarm is triggered.
Differential temperature between set point and upper
temperature alarm limit
Differential temperature between setpoint and lower
temperature alarm limit
As soon as a temperature alarm limit of "2" is selected, the
hysteresis is automatically reduced to 1
Corresponds to the temperature range: -10 °C to +50 °C
(4 – 20 mA: at 5 °C = 8 mA)
This value will not be deleted even after restarting
Value has no influence on the output power
Connect the sensor and confirm with the
Hold the sensor in water and confirm with the
Disconnect the sensor and confirm with the
cable break detection triggers in the middle between
this value and that of dry.
Check here the current value
Check here the current value
Check here the current value
Connect the multimeter and press or hold
until the display matches 1mA.
Connect the multimeter and press and hold
keys until the display matches 20 mA.
-key.
-key.
-key. The
or
-keys
or
-
43