M&C PSS Series Instruction Manual page 24

Gas conditioning unit
Hide thumbs Also See for PSS Series:
Table of Contents

Advertisement

Note
To calibrate an mA output, proceed as follows:
1. Enter the PIN (see chapter "11.2 PIN Entry").
2. Tap and hold the
3. The display shows the code entry.
The cooler should then be in a steady state at 5 °C [41 °F] (absolute value control) and provide one of the following
values:
8 mA (in case of 4 - 20 mA)
5 mA (in case of 0 - 20 mA)
The cable length is not limited, and the cable does not need to be shielded.
24
If a calibration error occurs and the mA output has been calibrated, the
limiting values also change!
-key for a short moment.
Enter the code "310" for the calibration of the lower value of the mA output.
Now connect a current meter to the connector of the mA output. This current
meter should measure a value close to 1 mA. You can now adjust this value in
0.0054 mA steps with the up and down arrow keys. After the adjustment, the
current meter should display 1 mA as accurately as possible. Accept the value
with the
-key.
Enter the code "C311" for the calibration of the upper value of the mA output.
Now connect a current meter to the connector of the mA output. This current
meter should be used to measure a value close to 20 mA. You can now adjust
this value in 0.0054 mA steps with the up and down arrow keys until the
current meter displays 20 mA as accurately as possible. Accept the value with
the
-key.
SS5C, SS5C/2, SS5C/3 | 1.00.00