Option "heated sample line":
For the electrical supply of a heated line with PT-100 sensor a 7-pin connector is available.
For the electrical supply of other heated components (e.g., heated sample gas probe or heated filter) a 4-pole round
plug connector is available at the heated lines Part No 01B4050 or 01B4040. The maximum connected load is 6 A,
1380 W for the 230 V sample gas conditioning unit and 6 A, 690 W for the 115 V version.
The maximum length of a possible heated sample line is calculated as follows:
Max. connection power [W] - Power consumption of heated components (i.e. sample probe) [W]
L [m] =
Power consumption of heated sample line [W/m]
Option "analogue output of the sample gas cooler temperature, Part No 01G9010":
The connection of the mA signal output is a 4-pole circular connector (female).
The mA output signal is available at pin 1 + and 3 -.
Measuring range: -10 to 50 °C [14 to 122 °F], mA output 4 to 20 mA or 0 to 20 mA, factory setting 4-20 mA. To
select the mA output range and to calibrate the mA output see chapter 13.1 and 13.2.
The load is 500 Ohm.