14.3
LA CALIBRATION
The liquid alarm sensors LA1S with cable break detection installed by M&C are factory activated and calibrated. If
required, the basic accuracy of the liquid alarm sensors can be optimized by recalibration. Later-installed liquid
sensors must be activated and calibrated.
Note
To calibrate a liquid alarm sensor, proceed as follows:
1. Enter the PIN (see chapter "11.2 PIN Entry").
2. Tap and hold the
3. The display shows the code entry.
15

RECEIPT OF GOODS AND STORAGE

The stationary gas conditioning units SS5C, SS5C/2 and SS5C/3 are completely pre-installed units.
Immediately after arrival take the gas conditioning system and possible special accessories carefully out of
the packaging material.
Compare the goods with the items listed on the delivery note.
Check the goods for any damage caused during delivery and, if necessary, notify your transport insurance
company without delay of any damage discovered.
Note
www.mc-techgroup.com
Make sure that the LA is activated. If a retrofitted LA1S is not activated, the
calibration has no effect and will be discarded.
-key for a short moment.
First calibrate the "dry state" of the liquid alarm sensor.
To do this, leave the LA in the dry state and enter the code "210". Confirm the
displayed value with the
Then calibrate the "wet state" of the liquid alarm sensor.
Immerse the LA in the process-dependent condensate to calibrate the 100 %
wet condition via code "211". Confirm the displayed value with the
The equipment should be stored in a protected, frost-free room!
SS5C, SS5C/2, SS5C/3 | 1.00.00
-key.
-key.