Electrical Connections; Figure 11 Electrical Connections For 230 V - M&C PSS Series Instruction Manual

Gas conditioning unit
Hide thumbs Also See for PSS Series:
Table of Contents

Advertisement

17.2

ELECTRICAL CONNECTIONS

Warning
Note
The SS5C, SS5C/2 and SS5C/3 gas conditioning systems are available with either 230 V/50 Hz or with 115 V/60 Hz
(for circuit diagram see Appendix). The device is protected by a 4 A fuse as standard. The fuse is located on the
terminal mounting rail
1
If a temperature controller with heated gas sample line is used as an option, the fuse protection increases to 10 A.
The mains cable for the electrical connection is connected to the terminal positions 1, 2 and 3 of the
terminal mounting rail
Figure 11
Electrical connections for 230 V
30
Wrong supply voltage can damage the equipment. When connecting the
equipment, make sure that the supply voltage is identical with the
information provided on the model type plate!
For the erection of power installations with rated voltages up to 1000V,
the requirements of VDE 0100 and relevant standards and specifications
must be observed!
The main circuit is equipped with a fuse corresponding to the nominal
current (over current protection); for electrical details see technical data.
Use an external switch to disconnect the gas conditioning unit from the
power supply. An external switch suitable for the switching-on location
must be provided by the customer.
(see Figure 5).
1
(see Figure 5).
SS5C, SS5C/2, SS5C/3 | 1.00.00
www.mc-techgroup.com