20

MAINTENANCE

Observe the facility- and process-specific safety measures before carrying out maintenance work!
Warning
The frequency of the maintenance work depends on the operational process and can therefore only be
determined in each individual case. Maintenance instructions of the individual components can be found in the
instruction manuals for individual components.
All parts to be maintained are easily accessible and installed on the aluminium mounting plate of the sample gas
conditioning unit. These are (see Figure 5):
• The filter element of the pre-filter FP-2T 4;.
Note
• •Check the tube of the SR25.2-W
Note
• Check the diaphragm of the sample gas pump N3KPE (N5 KPE or N9 KPE optionally)
replace if necessary;
Note
www.mc-techgroup.com
Dangerous voltage!
It is necessary to take the equipment off the mains before any assembly,
maintenance or repair work is carried out.
In order to protect downstream analyzers, the wet filter element must
always be replaced in the event of a condensate breakthrough.
6
condensate pump every six months and replace if necessary.
The complete instruction manual of the SR25.2-W is available at
www.mc-techgroup.com.
The complete instruction manual of N3 KPE, N5 KPE and N9 KPE is
available at www.mc-techgroup.com.
SS5C, SS5C/2, SS5C/3 | 1.00.00
5
every six months and
35