Figure 6 Dimensions And Design: Ss5C/2 - M&C PSS Series Instruction Manual

Gas conditioning unit
Hide thumbs Also See for PSS Series:
Table of Contents

Advertisement

1
2
Terminal mounting rail
1
Cooler, sample gas IN
3
Diaphragm sample gas pump
5
Figure 6
Dimensions and design: SS5C/2
All components of the gas conditioning system are mounted on an aluminium plate and are freely accessible.
The gas cooler and an appropriate diaphragm gas pump
flow (for operating instructions for individual components, see Appendix). The possible combinations are
summarised in the following table:
Type of
cooler
SS5C
ECP1000C
SS5C/2
ECP2000C
SS5C/3
ECP3000C
When controlling the sample gas flow rate at the outlet of the conditioning unit (option "flow meter" or
downstream external flow meter), the minimum flow rates (see table) must be observed. If the required minimum
total flow rate is not reached, excessive overpressure can lead to premature destruction of the pump diaphragm.
www.mc-techgroup.com
Max. gas flow [Nl/h]
150
150
350
SS5C, SS5C/2, SS5C/3 | 1.00.00
4
3
6
Cooler, sample gas OUT
2
Fine filter FP-2T
4
Peristaltic pump SR25.2-W, condensate OUT
6
(all connections DN 4/6)
5
are installed depending on the required maximum gas
Min. gas flow [Nl/h]
60
60
200
5
6
Sample gas pump
N 3 KPE
N 3 KPE
N 9 KPE
15