Problem
Display
Cooler and
sample gas
pump are
Current cooler
running;
temperature(s)
Condensate in
the sample gas
line
23
PROPER DISPOSAL OF THE DEVICE
At the end of the service life of our products, it is important to take care of the appropriate disposal of obsolete
electrical and non-electrical devices. To help protect our environment, follow the rules and regulations of your
country regarding recycling and waste management.
38
Possible causes
Pump tube defect
Peristaltic pump SR25.2-
W does not work
Insufficient drying of
sample gas
Sensor has not turned
off pump.
SS5C, SS5C/2, SS5C/3 | 1.00.00
Check/Solution
Replace pump tube (see manual peristaltic pump SR25.2-
W);
ok?
Check peristaltic pump (see manual peristaltic pump
SR25.2-W);
ok?
Check ECPX000C (se ECPX000C cooler instruction
manual);
Check the LA sensor function