8

DESIGN SPECIFICATION

 Handle
 Peltier cooling unit
 Diaphragm sample gas pump N5-KPE
 Ring-pull
 Top of the case
 Heat exchanger
Figure 2
Design of the conditioning unit PSS-20, internal components
ll components of PSS-20 portable gas conditioning systems can be removed from the case. Open
A
the top and left panel with an inner hexagon M4 screwdriver, according to the maximum gas flow
rate discharged, install Peltier cooling unit Fig. 2/② and appropriate diaphragm pump Fig. 2/③
(Refer to the instruction manual of each component).
The minimum amount of flow is determined by the sample gas pump Fig. 2/③.If the required
minimum total flow rate is not reached, excessive overpressure can lead to premature
destruction of the pump diaphragm.
The gas cooler is equipped with a Duran glass heat exchanger
PVDF or stainless steel are optionally available.
12
PSS-20 | 1.00.01
Fig. 2/⑥
as standard. Heat exchangers in